• DocumentCode
    1059736
  • Title

    A calorimeter for optical fiber power standard and advantages in calibration of optical fiber power meters

  • Author

    Suzuki, Yasuyuki ; Chida, Naomichi ; Murata, Akihiro ; Inoue, Takemi

  • Author_Institution
    Yokogawa Electric Corp., Tokyo, Japan
  • Volume
    42
  • Issue
    2
  • fYear
    1993
  • fDate
    4/1/1993 12:00:00 AM
  • Firstpage
    430
  • Lastpage
    433
  • Abstract
    A calorimeter for optical fiber power measurement is developed. It has a compensative absorber to stabilize environmental disturbance, and two types of adapters for optical fibers and light beams. The uncertainty in measurements of 1-mW and 10-μW optical fiber power was evaluated as ±0.18% and ±0.45% respectively, expressed as two standard deviations. The calibration uncertainty of an optical fiber power meter using the calorimeter as a primary standard is reduced by more than one half compared to that of the conventional method using a laser beam calorimeter and photodiode-type transfer standard
  • Keywords
    calorimeters; compensation; measurement standards; optical fibre testing; power measurement; 1 mW; 10 muW; calibration; calorimeter; compensative absorber; environmental disturbance; optical fiber power standard; power meters; primary standard; standard deviations; uncertainty; Calibration; Optical fiber communication; Optical fibers; Optical pumping; Optical reflection; Optical sensors; Power measurement; Temperature dependence; Temperature sensors; Wavelength measurement;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.278598
  • Filename
    278598