Automatic RF voltage calibration with a primary voltage standard up to 1 GHz
Author
Gierke, Hartmut ; Grno, L. ; Janik, Dieter ; Munter, K.
Author_Institution
Phys.-Tech. Bundesanstalt, Braunschweig, Germany
Volume
42
Issue
2
fYear
1993
fDate
4/1/1993 12:00:00 AM
Firstpage
519
Lastpage
523
Abstract
An automatic system for the calibration of RF voltage transfer standards in the frequency range between 1 MHz and 1 GHz is described. It incorporates a coaxial dual dry-load calorimeter as a primary voltage standard, with an automated computer-controlled DC voltage balance. By means of comparison measurements with a thermal voltage converter, the estimated relative uncertainties (2σ) for frequencies between 1 MHz and 100 MHz have been reduced by a factor of two to values between 0.01% and 0.1%
Keywords
computerised instrumentation; measurement standards; microwave measurement; voltage measurement; 1 MHz to 1 GHz; 1 MHz to 100 MHz; RF voltage calibration; automatic system; coaxial dual dry-load calorimeter; computer-controlled DC voltage balance; primary voltage standard; relative uncertainties; thermal voltage converter; Bridge circuits; Calibration; Coaxial components; Frequency estimation; Frequency measurement; Measurement standards; Radio frequency; Resistors; Thin wall structures; Voltage;