• DocumentCode
    1059939
  • Title

    Automatic RF voltage calibration with a primary voltage standard up to 1 GHz

  • Author

    Gierke, Hartmut ; Grno, L. ; Janik, Dieter ; Munter, K.

  • Author_Institution
    Phys.-Tech. Bundesanstalt, Braunschweig, Germany
  • Volume
    42
  • Issue
    2
  • fYear
    1993
  • fDate
    4/1/1993 12:00:00 AM
  • Firstpage
    519
  • Lastpage
    523
  • Abstract
    An automatic system for the calibration of RF voltage transfer standards in the frequency range between 1 MHz and 1 GHz is described. It incorporates a coaxial dual dry-load calorimeter as a primary voltage standard, with an automated computer-controlled DC voltage balance. By means of comparison measurements with a thermal voltage converter, the estimated relative uncertainties (2σ) for frequencies between 1 MHz and 100 MHz have been reduced by a factor of two to values between 0.01% and 0.1%
  • Keywords
    computerised instrumentation; measurement standards; microwave measurement; voltage measurement; 1 MHz to 1 GHz; 1 MHz to 100 MHz; RF voltage calibration; automatic system; coaxial dual dry-load calorimeter; computer-controlled DC voltage balance; primary voltage standard; relative uncertainties; thermal voltage converter; Bridge circuits; Calibration; Coaxial components; Frequency estimation; Frequency measurement; Measurement standards; Radio frequency; Resistors; Thin wall structures; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.278615
  • Filename
    278615