• DocumentCode
    1060174
  • Title

    Intercomparison of NIST, NPL, PTB, and VSL thermal voltage converters from 100 kHz to 1 MHz

  • Author

    Kinard, J.R. ; Knight, R.B.D. ; Martin, P. ; Klonz, M. ; de Vreede, J.P.M. ; Dessens, J.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
  • Volume
    42
  • Issue
    2
  • fYear
    1993
  • fDate
    4/1/1993 12:00:00 AM
  • Firstpage
    615
  • Lastpage
    617
  • Abstract
    Coaxial, thermal voltage converters were hand-carried among NIST, NPL, PTB, and VSL for comparison of AC-DC difference from 100 kHz to 1 MHz. The highly varied methods and underlying principles on which AC-DC difference determinations are based in each laboratory, the transport standards used, and the results of the comparisons are presented. The AC-DC differences reported by the participating laboratories are in very good agreement. It is concluded that representative values of AC-DC difference in the 100 kHz to 1 MHz frequency range are established with 2σ uncertainties of 4-8 ppm
  • Keywords
    convertors; measurement standards; voltage measurement; 100 kHz to 1 MHz; AC-DC difference; NIST; NPL; PTB; VSL; laboratory; thermal voltage converters; transport standards; Analog-digital conversion; Coaxial components; Frequency conversion; Hardware; Laboratories; NIST; Resistors; Thermoelectricity; Uncertainty; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.278639
  • Filename