DocumentCode
1060174
Title
Intercomparison of NIST, NPL, PTB, and VSL thermal voltage converters from 100 kHz to 1 MHz
Author
Kinard, J.R. ; Knight, R.B.D. ; Martin, P. ; Klonz, M. ; de Vreede, J.P.M. ; Dessens, J.
Author_Institution
Nat. Inst. of Stand. & Technol., Gaithersburg, MD, USA
Volume
42
Issue
2
fYear
1993
fDate
4/1/1993 12:00:00 AM
Firstpage
615
Lastpage
617
Abstract
Coaxial, thermal voltage converters were hand-carried among NIST, NPL, PTB, and VSL for comparison of AC-DC difference from 100 kHz to 1 MHz. The highly varied methods and underlying principles on which AC-DC difference determinations are based in each laboratory, the transport standards used, and the results of the comparisons are presented. The AC-DC differences reported by the participating laboratories are in very good agreement. It is concluded that representative values of AC-DC difference in the 100 kHz to 1 MHz frequency range are established with 2σ uncertainties of 4-8 ppm
Keywords
convertors; measurement standards; voltage measurement; 100 kHz to 1 MHz; AC-DC difference; NIST; NPL; PTB; VSL; laboratory; thermal voltage converters; transport standards; Analog-digital conversion; Coaxial components; Frequency conversion; Hardware; Laboratories; NIST; Resistors; Thermoelectricity; Uncertainty; Voltage;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/19.278639
Filename
278639