DocumentCode :
1062403
Title :
Comparison of PMTs from three different manufacturers for the CMS-HF forward calorimeter
Author :
Akgun, U. ; Anderson, E.W. ; Ayan, A.S. ; Gülmez, E. ; Miller, M. ; Onel, Y. ; Schmidt, I. ; Winn, D.
Author_Institution :
Univ. of Iowa, Iowa City, IA, USA
Volume :
51
Issue :
4
fYear :
2004
Firstpage :
1909
Lastpage :
1915
Abstract :
The builders of the CMS forward hadron calorimeter established a set of specifications for readout PMTs that reflected the physics goals and mechanical needs of the CMS experiment. Three manufacturers, Hamamatsu, Photonis (Philips), and Electron Tubes (EMI), provided candidate PMTs based on these initial requirements. Timing, gain, dark current properties, and single photoelectron spectra of these candidate PMTs were measured. Results show that Hamamatsu PMTs (R7525HA) conform best to the specifications of the Hadron-Forward (HF) Calorimeter.
Keywords :
Cherenkov radiation; calorimeters; particle calorimetry; photoelectron spectra; photomultipliers; position sensitive particle detectors; CMS-HF forward hadron calorimeter; Cherenkov light; Electron Tubes PMT; Hamamatsu R7525HA PMT; Philips PMT; Photonis PMT; dark current properties; gain; high-energy physics experiment; photomultiplier tube; single photoelectron spectra; timing; Collision mitigation; Current measurement; Dark current; Electromagnetic interference; Electron tubes; Gain measurement; Hafnium; Manufacturing; Physics; Timing; CMS; CMS Hadron-Forward; Calorimetry; Cherenkov light; HF; PMT; SPES; calorimeter; high-energy physics experiment; photomultiplier tube; single photoelectron spectrum;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2004.832681
Filename :
1323238
Link To Document :