DocumentCode
1063318
Title
Development of a RadFET linear array for intracavitary in vivo dosimetry during external beam radiotherapy and brachytherapy
Author
Price, Robert A. ; Benson, Chris ; Joyce, Malcolm J. ; Rodgers, Kenneth
Author_Institution
Phys. Dept., Clatterbridge Centre for Oncology NHS Trust, UK
Volume
51
Issue
4
fYear
2004
Firstpage
1420
Lastpage
1426
Abstract
We present the details of a new linear array dosimeter consisting of a chain of semiconductors mounted on an ultra-thin (50 μm thick) flexible substrate and housed in an intracavitary catheter. The semiconductors, manufactured by NMRC Cork, have not been packaging and incorporate a passivation layer that allows them to be mounted on the substrate using flip-chip-bonding. This paper reports, for the first time, the construction of a multiple (ten) detector array suited to in vivo dosimetry in the rectum, esophagus and vagina during external beam radiotherapy, as well as being adaptable to in vivo dosimetry during brachytherapy and diagnostic radiology.
Keywords
MOSFET; biological effects of radiation; brachytherapy; dosimetry; patient diagnosis; semiconductor counters; 50 micron; MOSFET; NMRC Cork; RadFET linear array; brachytherapy; diagnostic radiology; esophagus; external beam radiotherapy; flip-chip-bonding; intracavitary catheter; intracavitary in vivo dosimetry; multiple detector array; passivation layer; rectum; semiconductor detectors; ultra-thin flexible substrate; vagina; Adaptive arrays; Brachytherapy; Catheters; Dosimetry; In vivo; Passivation; Semiconductor device manufacture; Semiconductor device packaging; Structural beams; Substrates;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2004.832570
Filename
1323707
Link To Document