Evaluation of a prototype small-animal PET detector with depth-of-interaction encoding
Author :
Burr, Kent C. ; Ivan, Adrian ; Castleberry, Donald E. ; LeBlanc, James W. ; Shah, Kanai S. ; Farrell, Richard
Author_Institution :
Gen. Electr. Res., Niskayuna, NY, USA
Volume :
51
Issue :
4
fYear :
2004
Firstpage :
1791
Lastpage :
1798
Abstract :
We report on the evaluation of a prototype small-animal positron emission tomography (PET) detector with depth-of-interaction (DOI) encoding. The detector consisted of an 8×8 array of scintillator crystals that were read out on the top and bottom by position sensitive avalanche photodiodes (PSAPD). Each scintillator crystal had dimensions of 1.65 mm×1.65 mm×22.00 mm, and the array had a pitch of 1.75 mm. The 14×4 mm PSAPDs were coupled to compact, discrete transimpedance preamplifiers. The DOI response was measured by illuminating the detector module from the side with an electronically collimated fan beam of 511 keV gamma rays. The signals from the corner contacts of the PSAPDs were used to identify the crystal of interaction, and the ratio of the total signals from the two PSAPDs was used to calculate the depth z at which the interaction took place. By stepping the crystal array in the z direction (perpendicular to the fan beam), we were able to determine the DOI resolution for each individual crystal. For measurements made at 10°C, the average DOI resolution was better than 3 mm. Energy resolution and timing resolution (versus a fast plastic scintillator coupled to a photomultiplier tube) were made under the same operating conditions. The average energy resolution across the array was 16.6% at 511 keV, and the average timing resolution was 3.9 ns. Importantly, the detector performance was maintained all the way out to the crystals at the edges of the PSAPDs.
Keywords :
avalanche photodiodes; gamma-ray detection; gamma-ray effects; nuclear electronics; photomultipliers; position sensitive particle detectors; positron emission tomography; readout electronics; solid scintillation detectors; 1.65 mm; 10 C; 14 mm; 22.00 mm; 511 keV; APD; PET; PSAPD; crystal array; depth-of-interaction encoding; detector module; detector performance; discrete transimpedance preamplifiers; electronically collimated fan beam; energy resolution; fast plastic scintillator; gamma rays; photomultiplier tube; position sensitive avalanche photodiodes; position sensitive detector; prototype small-animal positron emission tomography detector; read out; scintillator crystals; timing resolution; Avalanche photodiodes; Crystals; Encoding; Energy resolution; Position sensitive particle detectors; Positron emission tomography; Prototypes; Sensor arrays; Signal processing; Timing;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2004.832905
Filename :
1323769
Link To Document :
بازگشت