DocumentCode :
1064483
Title :
New 9 ,\\times, 9 and 10 ,\\times, 10 BGO Block Detector for Human PET Using PMT Quadrant Sharing D
Author :
Liu, Shitao ; Liu, Jiguo ; Zhang, Yuxuan ; Kim, Soonseok ; Li, Hongdi ; Ramirez, Rocio ; Wang, Yu ; Baghaei, Hossain ; Wong, Wai-Hoi
Author_Institution :
Univ. of Texas, Houston
Volume :
55
Issue :
1
fYear :
2008
Firstpage :
457
Lastpage :
462
Abstract :
Our objectives were to develop two 9times9 and 10times10 array high resolution position sensitive block detectors using low cost photomultiplier-quadrant-sharing (PQS) technology. These two blocks can decode 4.3 mm or 3.9 mm BGO crystals with 39 mm size regular round PMT. By using ESR film as reflector between crystals, we achieved 96% packing fractions for these two blocks. Compared with conventional block design using 19 mm or 26 mm PMT, Phi39 mm size PMT can reduce the number of PMT in a camera by 55%-75%. Use of BGO crystal can reduce crystal cost by more than 70% compared to LSO or GSO crystals that are used in some of the newest PET cameras. The BGO crystal sizes-4.3 mm or 3.9 mm-in these two PQS blocks were similar to or smaller than BGO, LSO, or GSO crystals (4-6.3 mm) in commercial human PET cameras. The spatial resolution of the PQS blocks was expected to be similar or better. List mode data of these two blocks were acquired with Na- 22 source. Crystal-decoding map of blocks and individual crystal spectra were derived. All 9times9 and 10times10 crystals were clearly decoded on the block decoding map. These two PQS detectors decoded 81 or 100 BGO crystals per PMT. The average peak to valley ratio of decoding map was 3.4 and 1.8 respectively. The light collection efficiency for all crystals in the blocks were 71%-100% and 66%-100% respectively. The average energy resolutions of all crystals were 16.8% and 17.2% respectively. Our data suggest that even at lower cost the new PQS PET camera may outperform recent conventional PET cameras.
Keywords :
biomedical equipment; image resolution; photomultipliers; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; BGO block detector; ESR film; PMT quadrant sharing design; block decoding map; crystal-decoding map; high-resolution position sensitive detectors; human PET cameras; light collection efficiency; photomultiplier; spatial resolution; Cameras; Costs; Crystals; Decoding; Humans; Optical films; Paramagnetic resonance; Position sensitive particle detectors; Positron emission tomography; Sensor arrays; Photomultiplier -quadrant-sharing (PQS); position sensitive detector; positron emission tomography (PET); scintillation detector;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.2007.912879
Filename :
4448502
Link To Document :
بازگشت