• DocumentCode
    1064992
  • Title

    Normalisation of Histogrammed List Mode Data

  • Author

    Thielemans, Kris ; Morel, Christian ; Jacobson, Matthew W. ; Kaempf, Jerome H. ; Mustafovic, Sanida

  • Author_Institution
    Hammersmith Imanet Ltd., London
  • Volume
    55
  • Issue
    1
  • fYear
    2008
  • Firstpage
    543
  • Lastpage
    551
  • Abstract
    Many PET scanners nowadays have the possibility to record event-by-event information, known as list mode data. This has the advantage of keeping the data in the highest possible resolution (both temporal and spatial). In most cases, list mode data are then binned into sinogram format before reconstruction. In this paper, we discuss at which stage normalisation factors should be introduced. It is shown that noise is greatly reduced by performing the normalisation after the binning. We illustrate this with acquired and simulated data for the quad HiDAC camera.
  • Keywords
    image scanners; medical image processing; positron emission tomography; PET scanners; quad HiDAC camera; sinogram format; stage normalisation factors; Animals; Cameras; Detectors; Event detection; Jacobian matrices; Medical services; Noise reduction; Positron emission tomography; Spatial resolution; Statistical analysis; Data normalisation; PET; list mode;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2007.914207
  • Filename
    4448544