DocumentCode
1064992
Title
Normalisation of Histogrammed List Mode Data
Author
Thielemans, Kris ; Morel, Christian ; Jacobson, Matthew W. ; Kaempf, Jerome H. ; Mustafovic, Sanida
Author_Institution
Hammersmith Imanet Ltd., London
Volume
55
Issue
1
fYear
2008
Firstpage
543
Lastpage
551
Abstract
Many PET scanners nowadays have the possibility to record event-by-event information, known as list mode data. This has the advantage of keeping the data in the highest possible resolution (both temporal and spatial). In most cases, list mode data are then binned into sinogram format before reconstruction. In this paper, we discuss at which stage normalisation factors should be introduced. It is shown that noise is greatly reduced by performing the normalisation after the binning. We illustrate this with acquired and simulated data for the quad HiDAC camera.
Keywords
image scanners; medical image processing; positron emission tomography; PET scanners; quad HiDAC camera; sinogram format; stage normalisation factors; Animals; Cameras; Detectors; Event detection; Jacobian matrices; Medical services; Noise reduction; Positron emission tomography; Spatial resolution; Statistical analysis; Data normalisation; PET; list mode;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.2007.914207