DocumentCode
1068564
Title
Geometric and magnetic measurements of the gantry dipole for the HICAT medical accelerator
Author
Moritz, G. ; Pschorn, I. ; Klos, F. ; Weipert, M.
Author_Institution
GSI Gesellschaft fuer Schwerionenforschung, Darmstadt, Germany
Volume
14
Issue
2
fYear
2004
fDate
6/1/2004 12:00:00 AM
Firstpage
612
Lastpage
615
Abstract
Part of the HICAT medical accelerator, under construction at Heidelberg, Germany, is a huge gantry dipole. The magnet design was already reported at MT-17. This paper summarizes the main construction features. The geometric measurements are made with a laser tracker and include the determination of the iron core dimensions, the fiducialization and the alignment of the magnetic measurement equipment. Goal of the magnetic measurements was to shim the effective field boundary, to measure the integral field quality and the absolute integral field values. Dynamic field data in the central and in the fringe field are also reported.
Keywords
beam handling equipment; biomedical equipment; cancer; magnetic field measurement; radiation therapy; superconducting magnets; HICAT medical accelerator; dipole magnet; dynamic field data; effective field boundary; fiducialization; gantry dipole; geometric measurements; iron core; laser tracker; magnetic measurement; magnetic measurements; Acceleration; Coordinate measuring machines; Integral equations; Iron; Lamination; Magnetic cores; Magnetic field measurement; Magnetic separation; Magnetic variables measurement; Saturation magnetization; Dipole magnet; fiducialization; gantry; laser tracker;
fLanguage
English
Journal_Title
Applied Superconductivity, IEEE Transactions on
Publisher
ieee
ISSN
1051-8223
Type
jour
DOI
10.1109/TASC.2004.830000
Filename
1324868
Link To Document