DocumentCode
1071100
Title
Small cryostem for operation of Visible Light Photon Counters (VLPC)
Author
Baumbaugh, B. ; Bishop, J. ; Biswas, N. ; Coney, L. ; Marchant, J. ; Masterson, E. ; Ruchti, R. ; Stavish, T. ; Warchol, J. ; Wayne, M.
Author_Institution
Dept. of Phys., Notre Dame Univ., IN, USA
Volume
43
Issue
3
fYear
1996
fDate
6/1/1996 12:00:00 AM
Firstpage
1741
Lastpage
1745
Abstract
We have designed, constructed, and operated a small cryostem which supports the operation of 32 channels of Visible Light Photon Counters (HISTE-IV VLPCs). The VLPCs are situated within a small enclosure which can be lowered into a 36-100 liter dewar. The enthalpy of the boil-off helium keeps the VLPCs cold, and allows the system to be operated with stability for many days within the desired temperature range of 6-7.5 K. The cryostem is instrumented with clear fiber waveguides which transport the light from an optical connector situated at the top of the cryostem and outside of the dewar to the photosensors at cryogenic temperatures within the dewar. Electrical signals from the VLPCs are amplified at room temperature using QPA02 preamplifiers. Details of design and performance are reviewed
Keywords
cryogenics; photodetectors; 6 to 7.5 K; HISTE-IV; He; QPA02 preamplifiers; Visible Light Photon Counters; cryostem; design; fiber waveguides; photosensors; Connectors; Counting circuits; Cryogenics; Helium; Instruments; Optical waveguides; Preamplifiers; Stability; Stimulated emission; Temperature distribution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/23.507214
Filename
507214
Link To Document