Current transient spectroscopy: A high-sensitivity DLTS system
Author
Borsuk, Judith A. ; Swanson, Richard M.
Author_Institution
Stanford University, Stanford, CA
Volume
27
Issue
12
fYear
1980
fDate
12/1/1980 12:00:00 AM
Firstpage
2217
Lastpage
2225
Abstract
A deep-level transient spectrometer (DLTS) has been built that measures current transients, rather than the usual capacitance transients, in p-n junctions. The system was developed to optimize sensitivity for the study of low-concentration processing-induced defects. An analysis of its performance as a function of timing parameters is presented. A noise analysis of both capacitance and current transient measurement is presented, showing that neither method has an inherent sensitivity advantage, and that both should be capable of detection limits of about 10-7to 10-8of the shallow doping concentration. Noise measurements indicate that a detection limit of 10-7times the doping has been obtained. Spectra of a processed p+-n diode with no intentional contamination show several defect levels in the 10-5ND , or 1010cm-3, range. Spectra of gold-doped p+-n diodes yield emission data in good agreement with accepted values.
Keywords
Capacitance measurement; Current measurement; Diodes; Doping; Noise measurement; P-n junctions; Performance analysis; Pollution measurement; Spectroscopy; Timing;