• DocumentCode
    1073699
  • Title

    Improved contrast ultrasound with tissue harmonic minimizing pulse

  • Author

    Krishnan, K.B. ; Thomenius, K.E.

  • Author_Institution
    GE India Technol. Centre, Bangalore
  • Volume
    55
  • Issue
    1
  • fYear
    2008
  • fDate
    1/1/2008 12:00:00 AM
  • Firstpage
    249
  • Lastpage
    253
  • Abstract
    We have computed predistorted pulses that minimize tissue harmonics to < -25 dB compared to fundamental over a 4-cm depth in the focal region using a Hankel transform-based algorithm. The results indicate their potential in achieving patient independent improvement in contrast with microbubbles using wideband ultrasonic transducers at high mechanical indices.
  • Keywords
    biological tissues; biomedical ultrasonics; ultrasonic transducers; Han-kel transform-based algorithm; mechanical indices; microbubbles; patient independent improvement; predistorted pulses; tissue harmonics; wideband ultrasonic transducers; Attenuation; Bandwidth; Focusing; Frequency; Pulse generation; Signal generators; Ultra wideband technology; Ultrasonic imaging; Ultrasonic transducer arrays; Ultrasonic transducers; Algorithms; Animals; Contrast Media; Humans; Image Enhancement; Image Interpretation, Computer-Assisted; Liver; Reproducibility of Results; Sensitivity and Specificity; Signal Processing, Computer-Assisted; Ultrasonography;
  • fLanguage
    English
  • Journal_Title
    Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0885-3010
  • Type

    jour

  • DOI
    10.1109/TUFFC.2008.634
  • Filename
    4454320