DocumentCode
1075245
Title
Comparison between beryllium-copper and tungsten high frequency air coplanar probes
Author
Carbonero, J.-L. ; Morin, Gdrard ; Cabon, Bdatrice
Author_Institution
Central R&D, SGS-Thomson Microelectron., Crolles, France
Volume
43
Issue
12
fYear
1995
fDate
12/1/1995 12:00:00 AM
Firstpage
2786
Lastpage
2793
Abstract
High frequency air coplanar probes using tungsten tips are now available for silicon wafer probing with aluminum pads. A comparative study of the beryllium-copper and tungsten behavior is presented in terms of contact resistance values, stability and reproducibility. The contact theory is summarized for practical purposes and probe cleaning methods are exposed. Finally, tungsten is demonstrated to be the best material for breaking the aluminum oxide over the pad to enable accurate high frequency probing
Keywords
UHF measurement; beryllium alloys; contact resistance; copper alloys; integrated circuit testing; microwave measurement; probes; stability; surface cleaning; test equipment; tungsten; Al pads; Be-Cu; GaAs; HF air coplanar probes; Si; Si wafer probing; W; W tips; contact resistance; contact theory; high frequency probing; probe cleaning methods; reproducibility; stability; Aluminum oxide; Cleaning; Contact resistance; Electrical resistance measurement; Frequency measurement; Hafnium; Probes; Silicon; Transistors; Tungsten;
fLanguage
English
Journal_Title
Microwave Theory and Techniques, IEEE Transactions on
Publisher
ieee
ISSN
0018-9480
Type
jour
DOI
10.1109/22.475636
Filename
475636
Link To Document