Title :
An Automatic Calibration System for Portable Radiation Monitors
Author :
García-Hernández, Jose Manuel ; Mondragón-Contreras, Luis ; Torres-Bribiesca, Marco Antonio ; López-Callejas, Regulo ; Ramírez-Jiménez, Francisco Javier ; Gutiérrez-Ocampo, Efren ; Vázquez-Cervantes, Raul Mario ; Cruz-Estrada, Pedro ; Peña-Eguiluz, Rosend
Author_Institution :
Dept. of Electron. Syst., Nat. Inst. of Nucl. Res. (ININ), Metepec
fDate :
6/1/2009 12:00:00 AM
Abstract :
A novel electromechanical system capable of 3D displacement has been designed and built to be used in the automatic calibration of portable radiation monitors. The relevant feature of the system is that can adjust very precisely the position of the monitor in front of the standard radiation sources. The distance between source and monitor can be adjusted from 40 cm to 700 cm, assuring a plusmn0.1 cm uncertainty. In order to ensure a reliable operation and comply with the incumbent official technical standards, a study of the power quality was conducted to prevent any form of undesirable operation as a result of electric disturbances.
Keywords :
calibration; nuclear electronics; radiation monitoring; automatic calibration system; electromechanical system; electronic instrumentation; portable radiation monitors; standard radiation sources; Calibration; Control systems; Electromechanical systems; Electronic equipment; Optimal control; Power quality; Power system reliability; Radiation detectors; Uncertainty; Velocity control; Automation; calibration; power quality; radiation detectors; velocity control;