Quenched and Non-Quenched Ar-Xe Penning Mixtures as Detection Media for a Gridded-Microstrip Gas Chamber X-Ray Detector
         
        
            Author : 
Carita, Liliana P M M ; Conde, Carlos A N ; Santos, Filomena P.
         
        
            Author_Institution : 
Dept. de Fis., Univ. de Coimbra, Coimbra
         
        
        
        
        
            fDate : 
6/1/2009 12:00:00 AM
         
        
        
        
            Abstract : 
The performance of quenched and non-quenched argon-xenon Penning mixtures in a Gridded-Microstrip Gas Chamber is studied for 5.9 keV X-rays. The best energy resolution obtained was 12.5% for the 95%Ar-5%Xe mixture. Charge multiplication gains one order of magnitude higher than in pure xenon were obtained for 95%Ar-5%Xe. With the addition of the quenching agent (methane), although the gas gain at the same anode voltage decreases, a better energy resolution of 12.3% is achieved for the 85.5%Ar-5%Xe-9.5%CH4 mixture.
         
        
            Keywords : 
X-ray detection; gas scintillation detectors; position sensitive particle detectors; anode voltage; charge multiplication gain; energy resolution; gas scintillation; gridded-microstrip gas chamber; nonquenched Ar-Xe penning mixtures; quenching; soft X-ray detection; Anodes; Cathodes; Electrons; Energy resolution; Fluctuations; Instruments; Microstrip; X-ray detection; X-ray detectors; X-ray imaging; Argon-xenon and P10-xenon gas mixtures; Penning mixtures; microstrip plate; x-ray gaseous detector;
         
        
        