Title
Noninvasive Control of the Power Transferred to an Implanted Device by an Ultrasonic Transcutaneous Energy Transfer Link
Author
Shmilovitz, D. ; Ozeri, Shaul ; Chua-Chin Wang ; Spivak, Boaz
Author_Institution
Sch. of Electr. Eng., Tel-Aviv Univ., Tel-Aviv, Israel
Volume
61
Issue
4
fYear
2014
fDate
Apr-14
Firstpage
995
Lastpage
1004
Abstract
Ultrasonic transcutaneous energy transfer is an effective method for powering implanted devices noninvasively. Nevertheless, the amount of power harvested by the implanted receiver is sensitive to the distance and orientation of the external transmitting transducer attached to the skin with respect to the implanted receiving transducer. This paper describes an ultrasonic power transfer link whose harvested power is controlled by an inductive link. A small (5 μF) storage capacitor voltage, which is part of the implanted unit, is allowed to swing between 3.8 and 3.5 V using hysteretic control. The two control states are indicated by excitation (while the implanted storage capacitor voltage decreases) or the absence of excitation of an implanted coil that is magnetically coupled to an external coil attached to the skin surface. A 35 mW Ultrasonic Transcutaneous Energy Transfer link was fabricated using two piezoelectric transducers of equal size (Fuji Ceramics C-2 PZT disc 15 mm × 3 mm) operated at a vibration frequency of 720 kHz. By applying the proposed hysteretic control, the captured power was effectively regulated for implantation depths of up to 85 mm.
Keywords
energy harvesting; piezoelectric transducers; prosthetic power supplies; skin; ultrasonic transducers; Fuji Ceramics C-2 PZT disc; capacitance 5 muF; external coil; external transmitting transducer; harvested power; hysteretic control; implanted coil; implanted device; implanted receiving transducer; inductive link; noninvasive control; piezoelectric transducers; power 35 mW; power transfer; size 15 mm; size 3 mm; skin; storage capacitor voltage; ultrasonic power transfer link; ultrasonic transcutaneous energy transfer link; voltage 3.5 V to 3.8 V; Acoustics; Band-pass filters; Capacitors; Coils; Implants; Transducers; Voltage control; Contactless power transfer; hysteresis control; implanted medical device; ultrasonic transcutaneous energy transfer; voltage regulation;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2013.2280460