Title :
A Gas-Actuated Anthropomorphic Prosthesis for Transhumeral Amputees
Author :
Fite, Kevin B. ; Withrow, Thomas J. ; Shen, Xiangrong ; Wait, Keith W. ; Mitchell, Jason E. ; Goldfarb, Michael
Author_Institution :
Dept. of Mech. & Aeronaut. Eng., Clarkson Univ., Potsdam, NY
Abstract :
This paper presents the design of a gas-actuated anthropomorphic arm prosthesis with 21 degrees of freedom and nine independent actuators. The prosthesis utilizes the monopropellant hydrogen peroxide as a gas generator to power nine pneumatic type actuators. Of the nine independent actuators, one provides direct- drive actuation of the elbow, three provide direct-drive actuation for the wrist, and the remaining five actuate an underactuated 17 degree of freedom hand. This paper describes the design of the prosthesis, including the design of small-scale high-performance servovalves, which enable the implementation of the monopropellant concept in a transhumeral prosthesis. Experimental results are given characterizing both the servovalve performance and the force and/or motion control of various joints under closed-loop control.
Keywords :
artificial limbs; closed loop systems; force control; motion control; pneumatic actuators; servomechanisms; valves; closed-loop control; force control; gas-actuated anthropomorphic arm prosthesis; monopropellant hydrogen peroxide; motion control; pneumatic type actuators; servovalve; transhumeral amputees; Actuators; Anthropomorphism; Elbow; Extremities; Humans; Joints; Motion control; Muscles; Neural prosthesis; Prosthetics; Monopropellant; myoelectric; pneumatic; prosthesis; transhumeral;
Journal_Title :
Robotics, IEEE Transactions on
DOI :
10.1109/TRO.2007.914845