• DocumentCode
    1082928
  • Title

    Ultrasound Current Source Density Imaging

  • Author

    Olafsson, Ragnar ; Witte, Russell S. ; Huang, Sheng-Wen ; Donnell, Matthew O.

  • Author_Institution
    Dept. of Biomed. Eng., Univ. of Michigan, Ann Arbor, MI
  • Volume
    55
  • Issue
    7
  • fYear
    2008
  • fDate
    7/1/2008 12:00:00 AM
  • Firstpage
    1840
  • Lastpage
    1848
  • Abstract
    Surgery to correct severe heart arrhythmias usually requires detailed maps of the cardiac activation wave prior to ablation. The pinpoint electrical mapping procedure is laborious and limited by its spatial resolution (5-10 mm). We propose ultrasound current source density imaging (UCSDI), a direct 3-D imaging technique that potentially facilitates existing mapping procedures with superior spatial resolution. The technique is based on a pressure-induced change in resistivity known as the acoustoelectric (AE) effect, which is spatially confined to the ultrasound focus. AE-modulated voltage recordings are used to map and reconstruct current densities. In this preliminary study, we tested UCSDI under controlled conditions and compared it with conventional electrical mapping techniques. A 2-D dipole field was produced by a pair of electrodes in a bath of 0.9% NaCl solution. Boundary electrodes detected the AE signal while a 7.5-MHz focused ultrasound transducer was scanned across the bath. UCSDI located the current source and sink to within 1 mm of their actual positions. A future UCSDI system potentially provides real-time 3-D images of the cardiac activation wave coregistered with anatomical ultrasound and would greatly facilitate corrective procedures for heart abnormalities.
  • Keywords
    acoustoelectric effects; bioacoustics; bioelectric phenomena; biomedical ultrasonics; cardiology; diseases; medical image processing; surgery; 3D imaging technique; acoustoelectric effect; boundary electrodes; cardiac activation wave; focused ultrasound transducer; heart abnormality; mapping procedures; pinpoint electrical mapping procedure; resistivity; severe heart arrhythmias; spatial resolution; surgery; ultrasound current source density imaging; Conductivity; Current density; Electrodes; Focusing; Heart; Image reconstruction; Spatial resolution; Surgery; Ultrasonic imaging; Voltage; Acoustics; acoustics; acoustoelectric effect; bioelectric phenomena; electrocardiography; Algorithms; Animals; Body Surface Potential Mapping; Heart Conduction System; Humans; Image Interpretation, Computer-Assisted; Imaging, Three-Dimensional; Ultrasonography;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2008.919115
  • Filename
    4457869