Condenser Shunt for Measurement of Higi-Frequency Currents of Large Magnitude
Author
Nyman, Alexander
Author_Institution
Consulting Engineer, Dubilier Condenser Corporation, New. York City
Volume
16
Issue
2
fYear
1928
Firstpage
208
Lastpage
217
Abstract
The necessity for an accurate ammeter for large high-frequency currents is pointed out. A new device consisting of a large condenser in parallel with a small condenser, and the latter carrying the current to a small thermo-couple ammeter, is described. A device of this nature can be made very accurate; in fact, comparable in accuracy to any available standards. The construction of the device includes provisions for reducing and restricting the electrostatic and electromagnetic field, due to large current, the reduction of distributed inductance and capacity, and a provision to prevent the resonance effect of high harmonics of the operating current. Provisions are also made for locating the measuring instrument at a distance from the circuit. Large ratings are possible by connecting a number of condenser units in parallel.
Keywords
Ammeters; Circuits; Current measurement; Electromagnetic fields; Electromagnetic measurements; Electrostatic measurements; Inductance; Instruments; Joining processes; Resonance;