• DocumentCode
    1093861
  • Title

    An FDTD model for calculation of gradient-induced eddy currents in MRI system

  • Author

    Liu, Feng ; Crozier, Stuart

  • Author_Institution
    Sch. of Inf. Technol. & Electr. Eng., Univ. of Queensland, Brisbane, Qld., Australia
  • Volume
    14
  • Issue
    3
  • fYear
    2004
  • Firstpage
    1983
  • Lastpage
    1989
  • Abstract
    In most magnetic resonance imaging (MRI) systems, pulsed magnetic gradient fields induce eddy currents in the conducting structures of the superconducting magnet. The eddy currents induced in structures within the cryostat are particularly problematic as they are characterized by long time constants by virtue of the low resistivity of the conductors. This paper presents a three-dimensional (3-D) finite-difference time-domain (FDTD) scheme in cylindrical coordinates for eddy-current calculation in conductors. This model is intended to be part of a complete FDTD model of an MRI system including all RF and low-frequency field generating units and electrical models of the patient. The singularity apparent in the governing equations is removed by using a series expansion method and the conductor-air boundary condition is handled using a variant of the surface impedance concept. The numerical difficulty due to the "asymmetry" of Maxwell equations for low-frequency eddy-current problems is circumvented by taking advantage of the known penetration behavior of the eddy-current fields. A perfectly matched layer absorbing boundary condition in 3-D cylindrical coordinates is also incorporated. The numerical method has been verified against analytical solutions for simple cases. Finally, the algorithm is illustrated by modeling a pulsed field gradient coil system within an MRI magnet system. The results demonstrate that the proposed FDTD scheme can be used to calculate large-scale eddy-current problems in materials with high conductivity at low frequencies.
  • Keywords
    biomedical MRI; eddy currents; finite difference time-domain analysis; superconducting magnets; MRI system; Maxwell equations; conductor-air boundary condition; finite-difference time-domain model; gradient-induced eddy currents; magnetic resonance imaging; noninvasive diagnosis; pulsed magnetic gradient fields; series expansion; superconducting magnet; surface impedance; Boundary conditions; Conductivity; Conductors; Eddy currents; Finite difference methods; Magnetic resonance imaging; Maxwell equations; Radio frequency; Superconducting magnets; Time domain analysis; Cylindrical coordinate; FDTD; MRI; eddy current; finite-difference time-domain; low frequency; magnetic resonance imaging; superconducting magnet;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/TASC.2004.830609
  • Filename
    1331462