  • Abstract
    Microwave displacement sensors are becoming increasingly popular as vital sign detectors and structural vibration detectors. Until now there is not any method known for calibrating them that does not require accessing the electronic board by inserting and probing test signals or does not require cumbersome equipment that is able to provide large linear displacements. Hence, proposed is a calibration method for I-Q correction that makes use only of an external mechanical oscillator (a metallic bar fixed at an edge). As a test device, a low-quality X-band transceiver with a strong imbalance has been used. The calibrated sensor is compared successfully with a high-performance seismic accelerometer.
  • Keywords
    calibration; displacement measurement; microwave detectors; microwave measurement; oscillators; I-Q imbalance correction; X-band transceiver; calibration method; electronic board; external mechanical oscillator; metallic bar; microwave displacement sensor; seismic accelerometer; structural vibration detector; vital sign detector;