    A beamformer for the acquisition of evoked potentials

  • Author

    McKinley, C.A. ; Parker, P.A.

  • Author_Institution
    New Brunswick Univ., Fredericton, NB, Canada
  • Volume
    38
  • Issue
    4
  • fYear
    1991
  • fDate
    4/1/1991 12:00:00 AM
  • Firstpage
    379
  • Lastpage
    382
  • Abstract
    Evoked potential (EP) measurements using a delay and sun beamformer acquisition system are described. This system has the potential for significantly improving the signal/noise ratio (SNR) in EP measurements. The performance of the beamformer is characterized in terms of the number of electrodes and cross-channel correlation. When compared to conventional ensemble averaging, the beamformer reduces the number of response repetitions required to achieve a given SNR by a factor which approaches the number of channels in the acquisition system.
  • Keywords
    bioelectric potentials; biomedical electronics; biomedical measurement; beamformer acquisition system; cross-channel correlation; electrodes; ensemble averaging; evoked potentials; Biomedical measurements; Delay; Electrodes; Finite impulse response filter; Noise generators; Phased arrays; Sensor arrays; Signal to noise ratio; Time measurement; Working environment noise; Electrodes; Evoked Potentials, Somatosensory; Fourier Analysis; Humans; Median Nerve; Signal Processing, Computer-Assisted;