    Performance Results of BGO Block Detectors Based on Flat Panel PS-PMT for PET

  • Author

    Liu, Huafeng ; Zhao, Qiang ; Shi, Pengcheng

  • Author_Institution
    Zhejiang Univ., Hangzhou
  • Volume
    54
  • Issue
    4
  • fYear
    2007
  • Firstpage
    1298
  • Lastpage
    1301
  • Abstract
    We have developed a new detector composed of a 20times20 array of 2times2times20 mm3 Bi4Ge3O12 (BGO) scintillators and a novel flat panel position sensitive photomultiplier tube(FP-PS-PMT) - Hamamatsu R8400-00-M256. The performance of the block detector was evaluated in terms of applicability to positron emission tomography (PET). The energy resolution was about 23% FWHM for 511 keV gamma-rays. The spatial resolution of the detector was evaluated by measuring the detector response functions (DRFs) for each BGO segment and by measuring the coincidence response functions (CRFs). The average FWHM values of the DRFs were 2.0 mm. For the CRFs performance, the average FWHM values were 2.82 mm at 0deg and 5.20 mm at 30deg.
  • Keywords
    biomedical equipment; bismuth compounds; coincidence techniques; gamma-ray detection; photomultipliers; position sensitive particle detectors; positron emission tomography; scintillation counters; BGO - System; BGO block detectors; coincidence response functions; detector response functions; detector spatial resolution; energy resolution; flat panel position sensitive photomultiplier tube; gamma-ray detectors; positron emission tomography; Anodes; Biomedical optical imaging; Detectors; Energy resolution; Laboratories; Optical films; Photomultipliers; Positron emission tomography; Sensor arrays; Spatial resolution; BGO block detector; PET; PS-PMT;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2007.901231
  • Filename
    4291749