DocumentCode :
1101120
Title :
Comparison of power deposition patterns produced by microwave and radio frequency cardiac ablation catheters
Author :
Lin, J.C. ; Wang, Yu-Long
Author_Institution :
Dept. of Electr. Eng., Illinois Univ., Chicago, IL
Volume :
30
Issue :
12
fYear :
1994
fDate :
6/9/1994 12:00:00 AM
Firstpage :
922
Lastpage :
923
Abstract :
Transcatheter microwave and radio frequency (RF) electrodes are used for the ablative treatment of cardiac arrhythmias. The authors compare the power deposition patterns of microwave antennas and RF electrodes in saline phantoms of biological material. The decrease in power deposition as a function of distance away from the electrode is nearly exponential for RF and is considerably steeper than microwave antennas. These results suggest that microwave antennas are capable of heating a larger volume and thus creating a greater lesion than RF
Keywords :
biomedical equipment; biothermics; cardiology; electrodes; microwave antennas; radiation therapy; radiofrequency heating; RF electrodes; ablative treatment; biological material; cardiac arrhythmias; exponential dependence; microwave antennas; microwave cardiac ablation catheters; power deposition; power deposition patterns; radio frequency cardiac ablation catheters; saline phantoms; transcatheter microwave electrodes;