DocumentCode
1101760
Title
Two-mode multiline laser stabilization using a Fabry-Perot filter under analog and digital computer control
Author
DeSerio, R. ; Ruff, George A. ; Wing, William H.
Author_Institution
University of Tennessee, Knoxville, TN, USA
Volume
20
Issue
2
fYear
1984
fDate
2/1/1984 12:00:00 AM
Firstpage
140
Lastpage
148
Abstract
A scanning Fabry-Perot interferometer, servolocked to one of the output frequencies of a multiline laser, is used as a tracking filter so that a second servoloop can accurately stabilize that laser line to features of its intensity-versus-frequency profile. More stable locks occur where two longitudinal modes lase simultaneously than where the single-mode line intensity is maximized. Both analog and digital servoloops have been used for stabilizing a continuous-wave CO laser cavity. The microcomputer-aided digital stabilization yields the more reliable frequency locks and needs negligible frequency dithering. It has produced short-term (<1s) laser linewidths less than 100 kHz and long-term (>100 s) instability estimated as 80 kHz at a laser operating frequency of 55 THz in a noisy laboratory environment.
Keywords
Carbon monoxide lasers; Fabry-Perot interferometers; Laser stability; Analog computers; Digital control; Digital filters; Fabry-Perot; Frequency; Laser modes; Laser stability; Laser theory; Optical control; Physics;
fLanguage
English
Journal_Title
Quantum Electronics, IEEE Journal of
Publisher
ieee
ISSN
0018-9197
Type
jour
DOI
10.1109/JQE.1984.1072363
Filename
1072363
Link To Document