Title :
A method to determine the calorimetric equivalence correction for a coaxial microwave microcalorimeter
Author_Institution :
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
fDate :
6/1/1994 12:00:00 AM
Abstract :
A way is presented to obtain the microcalorimeter correction factor by direct measurement rather than by an indirect estimate or modeling. The microcalorimeter is used to measure the effective efficiency of a reference standard thermistor mount. The correction factor accounts for the different thermal paths and losses in the microcalorimeter reference standard combination. The uncertainty in the measurement depends primarily on an accurate determination of the correction factor. This has been an especially difficult problem in the coaxial case because of the center conductor. The method requires the fabrication of components that duplicate the thermal and RF loss in the microcalorimeter and reference standard. Using the technique with the new National Institute of Standards and Technology (NIST) type N coaxial microcalorimeter has substantially reduced the systematic uncertainty. The total uncertainty is about one-half the uncertainty of the prior NIST standard at frequencies above 1 GHz
Keywords :
calorimeters; calorimetry; error compensation; heat losses; measurement errors; measurement standards; microwave measurement; power measurement; thermistors; 1 GHz; NIST; National Institute of Standards and Technology; RF loss; calorimetric equivalence correction; coaxial microwave microcalorimeter; correction factor; fabrication; indirect estimate; losses; microcalorimeter correction factor; microcalorimeter reference standard; modeling; reference standard thermistor mount; thermal paths; uncertainty; Bolometers; Coaxial components; Electromagnetic heating; Microwave theory and techniques; NIST; Radio frequency; Temperature; Thermistors; Uncertainty; Voltage;
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on