DocumentCode :
1120335
Title :
On making accurate measurements with a harmonic distortion meter
Author :
Macdonald, J.
Author_Institution :
Texas Instruments Inc., Dallas, Texas
Issue :
6
fYear :
1957
Firstpage :
160
Lastpage :
161
Abstract :
It is pointed out that an avoidable error occurs in making total harmonic distortion measurements by the usually recommended method with art instrument incorporating art average-responding output meter. The error arises from the fact that the average value of a full-wave rectified distortion signal can be reduced by adding a small phase-shifted fundamental component to the distortion. The error is always in the direction to make equipment seem better than it actually is and can amount to ten per cent or more. It can be readily avoided by changing the instrument operating procedure.
Keywords :
Distortion measurement; Harmonic distortion; Instruments; Phase distortion; Total harmonic distortion;
fLanguage :
English