DocumentCode :
1122923
Title :
Use of High Resolution DAQ System to Aid Diagnosis of HD2b, a High Performance {\\rm Nb}_{3}{\\rm Sn} Dipole
Author :
Lizarazo, Juan ; Doering, Dionisio ; Doolittle, Lawrence ; Galvin, James ; Caspi, Shlomo ; Dietderich, Daniel R. ; Felice, Helene ; Ferracin, Paolo ; Godeke, Arno ; Joseph, J.M. ; Lietzke, A.F. ; Ratti, Alessandro ; Sabbi, GianLuca ; Trillaud, Frederic ;
Author_Institution :
Lawrence Berkeley Nat. Lab., Berkeley, CA, USA
Volume :
19
Issue :
3
fYear :
2009
fDate :
6/1/2009 12:00:00 AM
Firstpage :
2345
Lastpage :
2349
Abstract :
A novel voltage monitoring system to record voltage transients in superconducting magnets is being developed at LBNL . This system has 160 monitoring channels capable of measuring differential voltages of up to 1.5 kV with 100 kHz bandwidth and 500 kS/s digitizing rate. This paper presents analysis results from data taken with a 16 channel prototype system. From that analysis we were able to diagnose a change in the current-temperature margin of the superconducting cable by analysing Flux-Jump data collected after a magnet energy extraction failure during testing of a high field Nb3Sn dipole.
Keywords :
niobium alloys; superconducting cables; superconducting magnets; tin alloys; type II superconductors; channel prototype system; current-temperature margin; digitizing rate; flux-jump data; high field dipole magnet; magnet energy extraction failure; superconducting cable; superconducting magnets; voltage monitoring system; voltage transients; ${rm Nb}_{3}{rm Sn}$; HD2; superconducting magnets;
fLanguage :
English
Journal_Title :
Applied Superconductivity, IEEE Transactions on
Publisher :
ieee
ISSN :
1051-8223
Type :
jour
DOI :
10.1109/TASC.2009.2019036
Filename :
5153110
Link To Document :
بازگشت