• DocumentCode
    1123730
  • Title

    Real-Time Fetal Heart Monitoring in Biomagnetic Measurements Using Adaptive Real-Time ICA

  • Author

    Waldert, Stephan ; Bensch, Michael ; Bogdan, Martin ; Rosenstiel, Wolfgang ; Schölkopf, Bernhard ; Lowery, Curtis L. ; Eswaran, Hari ; Preissl, Hubert

  • Author_Institution
    Univ. of Tubingen, Tubingen
  • Volume
    54
  • Issue
    10
  • fYear
    2007
  • Firstpage
    1867
  • Lastpage
    1874
  • Abstract
    Electrophysiological signals of the developing fetal brain and heart can be investigated by fetal magnetoencephalography (fMEG). During such investigations, the fetal heart activity and that of the mother should be monitored continuously to provide an important indication of current well-being. Due to physical constraints of an fMEG system, it is not possible to use clinically established heart monitors for this purpose. Considering this constraint, we developed a real-time heart monitoring system for biomagnetic measurements and showed its reliability and applicability in research and for clinical examinations. The developed system consists of real-time access to fMEG data, an algorithm based on independent component analysis (ICA), and a graphical user interface (GUI). The algorithm extracts the current fetal and maternal heart signal from a noisy and artifact-contaminated data stream in real-time and is able to adapt automatically to continuously varying environmental parameters. This algorithm has been named Adaptive Real-time ICA (ARICA) and is applicable to real-time artifact removal as well as to related blind signal separation problems.
  • Keywords
    cardiology; graphical user interfaces; independent component analysis; magnetoencephalography; patient monitoring; GUI; adaptive real-time independent component analysis; biomagnetic measurements; clinical examinations; environmental parameters; fMEG; fetal heart signal; fetal magnetoencephalography; graphical user interface; maternal heart signal; real-time fetal heart monitoring; Biology computing; Biomagnetics; Data mining; Fetal heart; Fetus; Graphical user interfaces; Independent component analysis; Magnetoencephalography; Real time systems; Working environment noise; Adaptive; clustering; fetal; independent component analysis; magnetoencephalography; real-time systems; Algorithms; Computer Systems; Diagnosis, Computer-Assisted; Fetal Monitoring; Humans; Magnetics; Magnetoencephalography; Principal Component Analysis; Reproducibility of Results; Sensitivity and Specificity;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2007.895749
  • Filename
    4303280