Title :
A readout ASIC for SPECT
Author :
Pettersen, D.M. ; Mikkelsen, S. ; Talebi, J. ; Meier, D.
Author_Institution :
IDEAS ASA, Fornebu, Norway
fDate :
6/1/2005 12:00:00 AM
Abstract :
A new low-noise multichannel application specific integrated circuit (ASIC) has been designed for use in nuclear medical imaging single photon emission computed tomography (SPECT), and prototypes have been produced and tested. The ASIC has 128 charge sensitive amplifiers (CSAs) followed by shapers, threshold discriminators, and sample-and-hold. The ASIC delivers analog signal amplitude and digital address from channels with signals above threshold, and optionally allows one to read signals from any channel using a random-access readout. The total power consumption is 358 mW. The ASIC parameters have been specified and measured as follows: each amplifier can measure charge within a linear dynamic range from -40000 e to +40000 e. The preamplifier noise is 56 e+11.6 e/pF at 10 μs shaping time. The ASIC is ideally suited for readout of radiation sensors with a few pF capacitance and less than 100 pA dc input current. The ASIC has been tested with silicon diode pad sensors: the diode had a capacitance of ≈1 pF and a leakage current of 20 pA. We measure an equivalent noise charge of 85 e, and achieve a triggering threshold of 800 e. Energy spectra from radioactive isotopes have been measured with energy resolution of 823 eV FWHM for 5.9 keV X-rays.
Keywords :
X-ray detection; application specific integrated circuits; biomedical imaging; nuclear electronics; preamplifiers; readout electronics; silicon radiation detectors; single photon emission computed tomography; 20 pA; 358 mW; SPECT; X-rays; analog signal amplitude; charge sensitive amplifiers; digital address; energy spectra; leakage current; low-noise multichannel application specific integrated circuit; nuclear medical imaging single photon emission computed tomography; preamplifier noise; radiation sensors; radioactive isotopes; random-access readout; readout ASIC; sample-and-hold; shapers; silicon diode pad sensors; threshold discriminators; triggering threshold; Application specific integrated circuits; Biomedical imaging; Capacitive sensors; Charge measurement; Circuit testing; Current measurement; Diodes; Electrical capacitance tomography; Prototypes; Single photon emission computed tomography; ASIC; SPECT; charge sensitive amplifiers; low noise; radiation detector; spectroscopy;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.2005.850933