Abstract :
An instrument has been constructed which simultaneously measures optical path length at two wavelengths and so permits determination of average refractive index over the common path and hence distance to 1 \\times 10^{-6} or better without extensive meteorological observations. Using a He-Ne laser (6328 Å) and a filtered high-pressure mercury arc lamp (3681 Å), the instrument is expected to measure distances to 15 kilometers with accuracies as high as a few parts in ten million. Preliminary measurements over a 5.3-km path give a standard deviation of 0.15 cm, or better than 3 \\times 10^{-7} .
Keywords :
Density measurement; Instruments; Length measurement; Measurement standards; Meteorology; Optical filters; Optical refraction; Optical variables control; Refractive index; Wavelength measurement;