• DocumentCode
    112637
  • Title

    A Novel System Identification Technique for Improved Wearable Hemodynamics Assessment

  • Author

    Wiens, Andrew D. ; Inan, Omer T.

  • Author_Institution
    Sch. of Electr. & Comput. Eng., Georgia Inst. of Technol., Atlanta, GA, USA
  • Volume
    62
  • Issue
    5
  • fYear
    2015
  • fDate
    May-15
  • Firstpage
    1345
  • Lastpage
    1354
  • Abstract
    Recent advances have led to renewed interest in ballistocardiography (BCG), a noninvasive measure of the small movements of the body due to cardiovascular events. A broad range of platforms have been developed and verified for BCG measurement including beds, chairs, and weighing scales: while the body is coupled to such a platform, the cardiogenic movements are measured. Wearable BCG, measured with an accelerometer affixed to the body, may enable continuous, or more regular, monitoring during the day; however, the signals from such wearable BCGs represent local or distal accelerations of skin and tissue rather than the whole body. In this paper, we propose a novel method to reconstruct the BCG measured with a weighing scale (WS BCG) from a wearable sensor via a training step to remove these local effects. Preliminary validation of this method was performed with 15 subjects: the wearable sensor was placed at three locations on the surface of the body while WS BCG measurements were recorded simultaneously. A regularized system identification approach was used to reconstruct the WS BCG from the wearable BCG. Preliminary results suggest that the relationship between local and central disturbances is highly dependent on both the individual and the location where the accelerometer is placed on the body and that these differences can be resolved via calibration to accurately measure changes in cardiac output and contractility from a wearable sensor. Such measurements could be highly effective, for example, for improved monitoring of heart failure patients at home.
  • Keywords
    accelerometers; biomedical equipment; calibration; cardiovascular system; electrocardiography; gait analysis; haemodynamics; medical signal processing; patient monitoring; signal reconstruction; skin; WS BCG measurements; WS BCG reconstruction; accelerometer; ballistocardiography; biological tissue; calibration; cardiogenic movements; cardiovascular events; distal accelerations; heart failure patient monitoring; noninvasive measurement; regularized system identification approach; skin; small body movements; wearable BCG; wearable hemodynamics; wearable sensor; Accelerometers; Biomedical monitoring; Electrocardiography; Finite impulse response filters; Skin; Wearable sensors; Ballistocardiography; ballistocardiography; home health monitoring; sensor informatics; wearable sensing;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2014.2387354
  • Filename
    7001053