• DocumentCode
    1128406
  • Title

    An optical cable for a CO2laser scalpel

  • Author

    Ishiwatari, Hiromasa ; Ikedo, Masaru ; Tateishi, Fumikazu

  • Author_Institution
    Matushita Electric Industrial Co., Ltd., Moriguchi, Osaka, Japan
  • Volume
    4
  • Issue
    8
  • fYear
    1986
  • fDate
    8/1/1986 12:00:00 AM
  • Firstpage
    1273
  • Lastpage
    1279
  • Abstract
    This paper describes requirements of designing an optical cable for a CO2laser scalpel and a design concept satisfying these requirements. The optical fiber used is polycrystalline KRS-5 and the power delivered by this optical fiber ranges from 20 to 40 W. Since the optical cable consists of three parts-a connector, a cable, and a hand-piece-these parts have been discussed separately from the mechanical and optical point of view. Safety considerations also have been described especially in the case of fiber breaking and melting down.
  • Keywords
    Biomedical equipment safety; Laser biomedical applications; Optical fiber cables; Biomedical optical imaging; Connectors; Fiber lasers; Gas lasers; Laser surgery; Optical design; Optical fiber cables; Optical fiber communication; Optical fiber testing; Optical fibers;
  • fLanguage
    English
  • Journal_Title
    Lightwave Technology, Journal of
  • Publisher
    ieee
  • ISSN
    0733-8724
  • Type

    jour

  • DOI
    10.1109/JLT.1986.1074867
  • Filename
    1074867