DocumentCode :
113145
Title :
Setup and Validation of an MPI Signal Chain for a Drive Field Frequency of 150 kHz
Author :
Sattel, Timo F. ; Woywode, Oliver ; Weizenecker, Jurgen ; Rahmer, Jurgen ; Gleich, Bernhard ; Borgert, Jorn
Author_Institution :
Philips Med. Syst. DMC GmbH, Hamburg, Germany
Volume :
51
Issue :
2
fYear :
2015
fDate :
Feb. 2015
Firstpage :
1
Lastpage :
3
Abstract :
In this paper, an imaging system is considered, which is part of an activity to test the feasibility of clinical (whole-body) magnetic particle imaging (MPI). Recent studies on nerve stimulation in humans indicate that drive field amplitudes have to be limited to lower values compared with preclinical MPI. Since nerve stimulation thresholds increase with frequency, the drive field frequency for the clinical demonstrator is set to 150 kHz. To date, available MPI systems usually apply drive fields at frequencies ranging from 1 to 25 kHz. We report on the technical feasibility of a signal chain setup for a drive field frequency of 150 kHz.
Keywords :
biomedical equipment; biomedical imaging; magnetic particles; MPI signal chain; clinical demonstrator; clinical magnetic particle imaging; drive field amplitudes; drive field frequency; frequency 150 kHz; imaging system; nerve stimulation thresholds; Band-pass filters; Capacitors; Coils; Damping; Harmonic analysis; Imaging; Magnetic particles; Magnetic particle imaging (MPI); signal chain;
fLanguage :
English
Journal_Title :
Magnetics, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9464
Type :
jour
DOI :
10.1109/TMAG.2014.2326256
Filename :
7067537
Link To Document :
بازگشت