DocumentCode
1132197
Title
Precision AC–DC Transfer Measurement System Based on a 1000-V Inductive Voltage Divider
Author
Budovsky, Ilya ; Hagen, Thomas
Author_Institution
Nat. Meas. Inst. of Australia, Lindfield, NSW
Volume
58
Issue
4
fYear
2009
fDate
4/1/2009 12:00:00 AM
Firstpage
844
Lastpage
847
Abstract
The traceability for thermal voltage converters (TVCs) usually depends on a range-to-range buildup technique that requires either very small voltage-level dependencies of the converters or precise knowledge of these level dependencies. In practice, this inevitably leads to uncertainties that accumulate in the buildup process and dominate the overall uncertainty of the high-voltage ac-dc transfer measurement. This paper presents an alternative technique that uses a 1000-V inductive voltage divider (IVD) developed at the national measurement institute, linfield, Australia (NMIA). The use of the IVD eliminates the need for the buildup, which leads to a significant reduction in the uncertainty. The standard uncertainty of the new high-voltage TVC calibration setup is on the order of 1 muV/V, which is significantly smaller than the uncertainties derived from the buildup technique.
Keywords
voltage dividers; voltage measurement; Australia; Linfleld; National Measurement Institute; high-voltage TVC calibration setup; inductive voltage divider; precision AC-DC transfer measurement system; range-to-range buildup technique; thermal voltage converters; voltage 1000 V; AC–DC transfer; inductive voltage divider (IVD); thermal voltage converter (TVC);
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2008.2011084
Filename
4768713
Link To Document