• DocumentCode
    1135708
  • Title

    Linear Superposition Electrical Impedance Tomography Imaging With Multiple Electrical/Biopsy Probes

  • Author

    Ivorra, Antoni ; Shini, Mohanad ; Rubinsky, Boris

  • Author_Institution
    Depts. of Mech. Eng. & Bioeng., Univ. of California at Berkeley, Berkeley, CA
  • Volume
    56
  • Issue
    5
  • fYear
    2009
  • fDate
    5/1/2009 12:00:00 AM
  • Firstpage
    1465
  • Lastpage
    1472
  • Abstract
    In medical diagnostics, tissue is often examined with multiple discrete biopsies taken under ultrasound placement. In a previous theoretical study, we have suggested that the linear nature of the equations used in electrical impedance tomography (EIT) can be employed with the conventional practice of biopsy sampling to produce an image of the tissue between the biopsy samplings. Specifically, the biopsy probes can be used to record EIT-type electrical data during the discrete tissue sampling. The location of the discrete biopsy needle insertions available from the ultrasound placement of the probes can be combined with the electrical measurement data and used with linear superposition to produce a complete EIT image of the tissue between the sampled sites. In this study, we explore the concept experimentally using gel phantoms to simulate tissue and heterogeneities in the tissue. The experiments are performed in 2-D and 3-D configurations, and data are taken discretely, one at a time, through single electrical probe insertions. In the 2-D configuration, we were able to produce images of reasonable quality for heterogeneities with a diameter larger than 3 mm (conductivity ratio 1:5) and with relative conductivity differences above 50% (diameter 5 mm).
  • Keywords
    biological tissues; electric impedance imaging; medical image processing; phantoms; 2D configuration; biological tissue; biopsy sampling; conductivity ratio; discrete tissue sampling; gel phantoms; linear superposition electrical impedance tomography imaging; medical diagnostics; multiple electrical/biopsy probes; single electrical probe insertion; ultrasound placement; Biopsy; Conductivity; Equations; Image sampling; Impedance; Medical diagnosis; Needles; Probes; Tomography; Ultrasonic imaging; Biopsy; electrical impedance tomography (EIT); minimally invasive medical imaging; Agar; Biopsy, Needle; Electric Impedance; Equipment Design; Humans; Image Interpretation, Computer-Assisted; Image Processing, Computer-Assisted; Phantoms, Imaging; Tomography; Ultrasonography;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2009.2013821
  • Filename
    4770182