DocumentCode :
1137538
Title :
Orderly stimulation of skeletal muscle motor units with tripolar nerve cuff electrode
Author :
Baratta, Richard ; Ichie, Masayoshi ; Hwang, Sung Kwan ; Solomonow, Moshe
Author_Institution :
Louisiana State Univ. Med. Center, New Orleans, LA, USA
Volume :
36
Issue :
8
fYear :
1989
Firstpage :
836
Lastpage :
843
Abstract :
An electrical nerve stimulation technique, using a single tripolar electrode, was shown to be capable of recruiting motor units according to their size, while allowing simultaneous but independent control of firing rate in the active units. Test paradigms consisting of established fundamental physiological concepts of soleus-gastrocnemius architecture, motor units conduction velocity, firing rate behavior of motor units of different sizes, and their susceptibility to fatigue were used to validate the technique and demonstrate its utility as a basic and applied research tool.
Keywords :
bioelectric phenomena; biological techniques and instruments; muscle; neurophysiology; biological research technique; electrical nerve stimulation technique; fatigue susceptibility; firing rate behavior; gastrocnemius; motor unit conduction velocity; muscle architecture; soleus; test paradigms; tripolar nerve cuff electrode; Animals; Electrodes; Fatigue; Frequency; In vivo; Muscles; Nerve fibers; Orthopedic surgery; Recruitment; Testing; Action Potentials; Animals; Cats; Electric Stimulation; Electrodes, Implanted; Electromyography; Equipment Design; Fatigue; Hindlimb; Motor Neurons; Muscle Contraction; Muscles; Neural Conduction; Reaction Time;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/10.30809
Filename :
30809
Link To Document :
بازگشت