• DocumentCode
    1138426
  • Title

    Neutron detection experiments with ionization counter and 6Li converter

  • Author

    Engels, R. ; Clemens, U. ; Kemmerling, G. ; Schelten, J.

  • Author_Institution
    Forschungszentrum Juelich GmbH, Germany
  • Volume
    51
  • Issue
    5
  • fYear
    2004
  • Firstpage
    2131
  • Lastpage
    2134
  • Abstract
    The detection of neutrons with thin 6Li containing films or layers within an ionization chamber is discussed. A detection efficiency of 50% for thermal neutrons is obtainable with three 100-μm thin 6Li foils. Preliminary measurements were performed with converters which contain a 1-μm thin 6LiF layer. The detector efficiency is only 3%. Pulse height distributions were measured and analyzed for the gases Ar, CF4, and C2F6 at pressures between 0.5 and 5 atm and at voltages between 50 and 500 V. Diffraction peaks from polycrystalline iron and semi-crystalline pyrolytic graphite were measured in TOF mode and are presented.
  • Keywords
    argon; convertors; graphite; ionisation chambers; iron; lithium; lithium compounds; neutron detection; organic compounds; pulse height analysers; scintillation counters; thin films; 0.5 to 5 atm; 1 micron; 100 micron; 50 to 500 V; 6Li converter; Ar; C; Fe; LiF; TOF mode; carbon tetrafluoride; detection efficiency; diffraction peaks; hexa fluoroethylene; ionization chamber; ionization counter; neutron detection experiments; polycrystalline iron; preliminary measurements; pulse height distributions; semicrystalline pyrolytic graphite; thermal neutrons; thin 6Li foils; thin 6LiF layer; Argon; Counting circuits; Detectors; Diffraction; Gases; Ionization chambers; Neutrons; Performance evaluation; Pulse measurements; Voltage;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2004.836040
  • Filename
    1344296