    Comparative Measurements on \\rm LaBr_3(\\rm Ce) and \\rm LaCl_3(\\rm Ce) Scintillators Coupled to PSPMT

  • Author

    Lee, Wonho ; Wehe, David K. ; Kim, Byoungjik

  • Author_Institution
    Radiat. Meas. & Imaging Group, Univ. of Michigan, Ann Arbor, MI, USA
  • Volume
    52
  • Issue
    4
  • fYear
    2005
  • Firstpage
    1119
  • Lastpage
    1122
  • Abstract
    Lanthanum halide detector energy resolution measurements at 662 keV for a  LaBr_3 crystal (3.67%) and a 6 \\times 6 array of  LaCl_3 crystals (4.65%) coupled to a 5 \\prime \\prime diameter Hamamatsu PSPMT are compared to measurements on arrays of NaI (7.37%) and BGO (16.07%). The uncorrected energy resolution using the entire  LaCl_3 array is poor (13%) due to weaker responses from the edge pixels, but is improved to 5.85% with a position dependent gain correction. Spectra are presented from a well-collimated source focused upon a small area of the detector, as well as uncollimated 2D flood-fields.
  • Keywords
    gamma-ray detection; lanthanum compounds; position sensitive particle detectors; solid scintillation detectors; 5 in; 662 keV; BGO detector comparison; LaBr3(Ce) scintillators; LaCl3(Ce) scintillators; NaI detector comparison; edge pixels; energy resolution measurements; gamma ray cameras; lanthanum halide detector; Crystals; Degradation; Detectors; Energy measurement; Energy resolution; Helium; Lanthanum; Linearity; Sensor arrays; Timing; lanthanum halide;