• DocumentCode
    1141242
  • Title

    Calibrating network analyzers with imperfect test ports

  • Author

    Juroshek, John R. ; Hoer, Cletus A. ; Kaiser, Raian F.

  • Author_Institution
    Nat. Inst. of Stand. & Technol., Boulder, CO, USA
  • Volume
    38
  • Issue
    4
  • fYear
    1989
  • fDate
    8/1/1989 12:00:00 AM
  • Firstpage
    898
  • Lastpage
    901
  • Abstract
    The test ports on automatic network analyzers are generally built with an impedance that matches the impedance of the calibration standards. Experimental evidence is presented that substantial impedance discontinuities can be tolerated at the test port interface if proper calibration procedures are observed. The 50-Ω test port on one of the six-ports in a dual six-port network analyzer was replaced with a 75-Ω test port. This test port was then calibrated to look like a 50-Ω test port. Measurements on various devices show that it is possible to make a 75-Ω test port indistinguishable from a 50-Ω test port
  • Keywords
    calibration; measurement standards; network analysers; 50 ohm; 75 ohm; automatic network analyzers; calibration standards; dual six-port network analyzer; impedance discontinuities; Automatic testing; Calibration; Coaxial components; Conductors; Impedance; Manufacturing; Measurement errors; Modems; Thumb; Transmission lines;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.31010
  • Filename
    31010