Calibrating network analyzers with imperfect test ports
Author
Juroshek, John R. ; Hoer, Cletus A. ; Kaiser, Raian F.
Author_Institution
Nat. Inst. of Stand. & Technol., Boulder, CO, USA
Volume
38
Issue
4
fYear
1989
fDate
8/1/1989 12:00:00 AM
Firstpage
898
Lastpage
901
Abstract
The test ports on automatic network analyzers are generally built with an impedance that matches the impedance of the calibration standards. Experimental evidence is presented that substantial impedance discontinuities can be tolerated at the test port interface if proper calibration procedures are observed. The 50-Ω test port on one of the six-ports in a dual six-port network analyzer was replaced with a 75-Ω test port. This test port was then calibrated to look like a 50-Ω test port. Measurements on various devices show that it is possible to make a 75-Ω test port indistinguishable from a 50-Ω test port
Keywords
calibration; measurement standards; network analysers; 50 ohm; 75 ohm; automatic network analyzers; calibration standards; dual six-port network analyzer; impedance discontinuities; Automatic testing; Calibration; Coaxial components; Conductors; Impedance; Manufacturing; Measurement errors; Modems; Thumb; Transmission lines;