• DocumentCode
    1142995
  • Title

    A Detailed Monte Carlo Study of Multiple Scattering Contamination in Compton Tomography at 90??

  • Author

    Guerra, Alberto Del ; Bellazzini, Ronaldo ; Tonelli, Guido ; Venturi, Renzo ; Nelson, Walter R.

  • Volume
    1
  • Issue
    2
  • fYear
    1982
  • Firstpage
    147
  • Lastpage
    152
  • Abstract
    A low dose technique has been recently proposed for tomographic studies of the lung, which makes use of a gamma camera to detect 90?? Compton-scattered photons from external planar gamma source. In this paper, we present a detailed Monte Carlo study of this technique. A 20 ?? 20 ?? 20 cm3 water phantom was simulated as a target and a large gamma camera equipped with an imaging collimator as a detector. The multiple scattering contamination of the single scattered signal was studied as a function of the source-detector geometry and of the incident energy in the range 100-500 keV. The multiple to single scatter ratio has an approximate 1/E0.7 dependence and increases almost linearly with the phantom depth and the transversal thickness at 90??. Simulation has been also performed with a 16 ?? 16 ?? 10 cm3 sawdust phantom of 0.3 g/cm3 density; the Monte Carlo results agree to within a few percent with experimental data.
  • Keywords
    Cameras; Contamination; Electromagnetic scattering; Gamma ray detection; Gamma ray detectors; Imaging phantoms; Lungs; Monte Carlo methods; Particle scattering; Tomography;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1982.4307562
  • Filename
    4307562