• DocumentCode
    1143046
  • Title

    Undersampling Errors in Region-of-Interest Tomography

  • Author

    Sankar, P.V. ; Nalcioglu, O. ; Sklansky, J.

  • Volume
    1
  • Issue
    3
  • fYear
    1982
  • Firstpage
    168
  • Lastpage
    173
  • Abstract
    A detailed error analysis for the dual-sampling region-ofinterest X-ray tomography is presented. Simulation studies are used along with a range of sampling rates to quantitate the amount of sampling errors within the region of interest. It is shown that as the rate of sampling outside the region of interest becomes sparse the amount of sampling errors within the region of interest increases considerably.
  • Keywords
    Computed tomography; Diseases; Error analysis; Image reconstruction; Image resolution; Image sampling; Interpolation; Sampling methods; Spatial resolution; X-ray tomography;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1982.4307568
  • Filename
    4307568