DocumentCode
1143503
Title
High sensitivity detectors for measurement of space potential fluctuations
Author
Keating, L.M. ; Schoch, P.M. ; Crowley, T.P. ; Russell, W.G. ; Schatz, J.G., Jr.
Author_Institution
Rensselaer Polytech. Inst., Troy, NY, USA
Volume
22
Issue
4
fYear
1994
fDate
8/1/1994 12:00:00 AM
Firstpage
424
Lastpage
429
Abstract
A detection technique yielding an order of magnitude increase in the sensitivity of a Proca-Green type energy analyzer to beam energy fluctuations has been demonstrated. A sensitivity of <1 part in 10 6 is possible. The technique can be used to increase the sensitivity of a Heavy Ion Beam Probe (HIBP) to space potential fluctuations in a plasma. The increase in sensitivity is achieved by splitting the detected ion beam into 10 smaller beamlets and summing the signals produced by the beamlets. Experimental results from an ion beam test facility are presented along with the projected results if implemented on the 2 MeV HIBP on the TEXT-Upgrade tokamak
Keywords
Tokamak devices; fluctuations; ion beams; plasma diagnostics; plasma probes; plasma toroidal confinement; sensitivity; space charge; Proca-Green type energy analyzer; TEXT-Upgrade tokamak; beam energy fluctuations; beamlets; detection technique; heavy ion beam probe; high sensitivity detectors; ion beam test facility; sensitivity; space potential fluctuations; splitting; Current measurement; Detectors; Electrostatic analysis; Energy measurement; Extraterrestrial measurements; Fluctuations; Geometry; Ion beams; Plasma measurements; Test facilities;
fLanguage
English
Journal_Title
Plasma Science, IEEE Transactions on
Publisher
ieee
ISSN
0093-3813
Type
jour
DOI
10.1109/27.310650
Filename
310650
Link To Document