  • Abstract
    The process of imaging by detection of the annihilation gamma rays generated from positron emitters which have been injected into a patient by a particle accelerator has been studied in detail. The relationships between patient dose and injected activity have been calculated for C-11, N-13, C-15, F-17, and Ne-19 and measured for C-11 and Ne-19 with good agreement with the calculations. The requirements for imaging of the small amounts of activity that can be injected safely have been analyzed in terms of one specific application of the radioactive beam injection technique, that of Bragg peak localization in support of radiotherapy by heavy ions. The characteristics of an existing camera with sufficient sensitivity and spatial accuracy for that task are described. Results of the calculations of radioactive beam flux requirements are shown.
  • Keywords
    Acceleration; Cameras; Gamma ray detection; Gamma ray detectors; Gamma rays; Image analysis; Linear particle accelerator; Optical imaging; Particle beams; Positrons;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1984.4307660
  • Filename