• DocumentCode
    1144129
  • Title

    Image Reconstruction from Cone-Beam Projections: Necessary and Sufficient Conditions and Reconstruction Methods

  • Author

    Smith, Bruce D.

  • Volume
    4
  • Issue
    1
  • fYear
    1985
  • fDate
    3/1/1985 12:00:00 AM
  • Firstpage
    14
  • Lastpage
    25
  • Abstract
    Previously unknown sufficient conditions, a necessary condition, and reconstruction methods for image reconstruction from cone-beam projections are developed. A sufficient condition developed is contained in the following statement. Statement 5: If one every plane that intersects the object, there exists at least one cone-beam source point, then the object can be reconstructed. Reconstruction methods for an arbitrary configuration of source points that satisfy Statement 5 are derived. By requiring additional conditions on the configuration of source points, a more efficient reconstruction method is developed. It is shown that when the configuration of source points is a circle, Statement 5 is not satisfied. In spite of this, several suggestions are made for reconstruction from a circle of source points.
  • Keywords
    Computed tomography; Engine cylinders; Heart; Image reconstruction; Optical imaging; Positron emission tomography; Radiation dosage; Reconstruction algorithms; Sufficient conditions; X-ray imaging;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1985.4307689
  • Filename
    4307689