• DocumentCode
    1144177
  • Title

    Effects of Detector Wobble Motion on Image Noise in Positron Emission Tomography

  • Author

    Palmer, M.R. ; Bergstrom, M. ; Beddoes, M.P. ; Pate, B.D.

  • Volume
    4
  • Issue
    1
  • fYear
    1985
  • fDate
    3/1/1985 12:00:00 AM
  • Firstpage
    58
  • Lastpage
    62
  • Abstract
    The effects of detector wobble motion on image variance is described analytically and verified by simulation studies. Simulation results in the form of variance images are presented for various phantoms, sampled under both uniform and wobble-motion conditions. It is shown that under most measurement conditions in a typical PET study, the image precision is degraded due to the nonuniform projection variance introduced by the wobble motion. Using the wobble parameters of the UBC-TRIUMF development of PETT VI, the rms noise is found to be increased by 9 percent due to wobbling.
  • Keywords
    Analysis of variance; Analytical models; Detectors; Image analysis; Image motion analysis; Imaging phantoms; Motion analysis; Motion detection; Motion measurement; Positron emission tomography;
  • fLanguage
    English
  • Journal_Title
    Medical Imaging, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0278-0062
  • Type

    jour

  • DOI
    10.1109/TMI.1985.4307694
  • Filename
    4307694