Abstract
The design and performance of spatial localization filters in an MR imaging system with time-varying gradients is presented. The system is based on recording measurements of the imaged volume-done in the presence of cosinusoidal gradients. A postprocessing scheme is shown which enables every point to be imaged from a single FID signal. Methods are shown which achieve a three-demensional localization with arbitrarily small sidelobes. Those localization filters are implemented without significant computational penalty. The signal-to-noise ratio (SNR) in the suggested imaging system is shown to be space-invariant (same SNR for each imaged point). We find a closed-form relation between the localization filter and the resulting noise. It is shown that the gain achieved by the localization filters highly exceeds the relatively small loss in SNR.
Keywords
Data acquisition; Filters; High-resolution imaging; Image resolution; Magnetic field measurement; Magnetic noise; Magnetic resonance imaging; Magnetic separation; Signal to noise ratio; Time varying systems;