• DocumentCode
    1144356
  • Title

    Resolution and Noise Considerations in MRI Systems with Time-Varying Gradients

  • Author

    Shenberg, Itzhak ; Macovski, Albert

  • Volume
    4
  • Issue
    3
  • fYear
    1985
  • Firstpage
    144
  • Lastpage
    152
  • Abstract
    The design and performance of spatial localization filters in an MR imaging system with time-varying gradients is presented. The system is based on recording measurements of the imaged volume-done in the presence of cosinusoidal gradients. A postprocessing scheme is shown which enables every point to be imaged from a single FID signal. Methods are shown which achieve a three-demensional localization with arbitrarily small sidelobes. Those localization filters are implemented without significant computational penalty. The signal-to-noise ratio (SNR) in the suggested imaging system is shown to be space-invariant (same SNR for each imaged point). We find a closed-form relation between the localization filter and the resulting noise. It is shown that the gain achieved by the localization filters highly exceeds the relatively small loss in SNR.
  • Keywords
    Data acquisition; Filters; High-resolution imaging; Image resolution; Magnetic field measurement; Magnetic noise; Magnetic resonance imaging; Magnetic separation; Signal to noise ratio; Time varying systems;