DocumentCode
1144430
Title
NMR Imaging for Magnets with Large Nonuniformities
Author
Sekihara, Kensuke ; Matsui, Shigeru ; kohno, Hideki
Volume
4
Issue
4
fYear
1985
Firstpage
193
Lastpage
199
Abstract
Nuclear magnet resonance (NMR) imaging in the presence of large static field nonuniformity is presented. The method involves three steps: field mapping, imaging by a conventional method, and image restoration from nonuniform field influence. Two-dimensional experiments are performed to demonstrate effectiveness of this method, as well as to clarify practical problems in employing the method. Experimental results show that, with this method, images free from distortion can be obtained using a magnet with an order of magnitude greater nonuniformity than magnets commonly used for whole-body imaging.
Keywords
Costs; Hospitals; Humans; Image restoration; Laboratories; Magnetic field measurement; Magnetic resonance imaging; Magnets; Nuclear magnetic resonance; Production;
fLanguage
English
Journal_Title
Medical Imaging, IEEE Transactions on
Publisher
ieee
ISSN
0278-0062
Type
jour
DOI
10.1109/TMI.1985.4307722