DocumentCode :
1144430
Title :
NMR Imaging for Magnets with Large Nonuniformities
Author :
Sekihara, Kensuke ; Matsui, Shigeru ; kohno, Hideki
Volume :
4
Issue :
4
fYear :
1985
Firstpage :
193
Lastpage :
199
Abstract :
Nuclear magnet resonance (NMR) imaging in the presence of large static field nonuniformity is presented. The method involves three steps: field mapping, imaging by a conventional method, and image restoration from nonuniform field influence. Two-dimensional experiments are performed to demonstrate effectiveness of this method, as well as to clarify practical problems in employing the method. Experimental results show that, with this method, images free from distortion can be obtained using a magnet with an order of magnitude greater nonuniformity than magnets commonly used for whole-body imaging.
Keywords :
Costs; Hospitals; Humans; Image restoration; Laboratories; Magnetic field measurement; Magnetic resonance imaging; Magnets; Nuclear magnetic resonance; Production;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.1985.4307722
Filename :
4307722
Link To Document :
بازگشت