DocumentCode :
1151893
Title :
Electronic Gating for Particle/Cell Counting and Sizing, DSP-operated
Author :
De Bisschop, Frank R E
Author_Institution :
Phys. Lab., Ghent Univ., Ghent, Belgium
Volume :
58
Issue :
9
fYear :
2009
Firstpage :
3159
Lastpage :
3166
Abstract :
A hardware and software solution is proposed for the development of a DSP-operated cell/particle counting and sizing system. LabVIEW, which is a graphical programming language, was adopted as the DSP programming platform. Special attention was paid to data reduction, noise elimination, baseline restoration, and pulsewidth discrimination. Global accuracy is verified based on externally generated test signals and using particle size standards. The hardware and software, with minor modifications, are also suitable as a detection system for cell/particle chromatography and flow cytometry.
Keywords :
biosensors; chromatography; digital signal processing chips; particle counting; particle size; DSP programming; DSP-operated cell-particle counting; LabVIEW; baseline restoration; cell-particle chromatography; data reduction; electronic gating; flow cytometry; graphical programming language; hardware-software solution; noise elimination; particle-cell counting; pulsewidth discrimination; Biomedical signal analysis; biomedical signal processing; biomedical transducers; cell/particle chromatography; cell/particle counting and sizing; flow cytometry;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2009.2017142
Filename :
5175329
Link To Document :
بازگشت