DocumentCode :
1153454
Title :
Magneto-Optical Tracking of Flexible Laparoscopic Ultrasound: Model-Based Online Detection and Correction of Magnetic Tracking Errors
Author :
Feuerstein, Marco ; Reichl, Tobias ; Vogel, Jakob ; Traub, Joerg ; Navab, Nassir
Author_Institution :
Dept. of Media Sci., Nagoya Univ., Nagoya
Volume :
28
Issue :
6
fYear :
2009
fDate :
6/1/2009 12:00:00 AM
Firstpage :
951
Lastpage :
967
Abstract :
Electromagnetic tracking is currently one of the most promising means of localizing flexible endoscopic instruments such as flexible laparoscopic ultrasound transducers. However, electromagnetic tracking is also susceptible to interference from ferromagnetic material, which distorts the magnetic field and leads to tracking errors. This paper presents new methods for real-time online detection and reduction of dynamic electromagnetic tracking errors when localizing a flexible laparoscopic ultrasound transducer. We use a hybrid tracking setup to combine optical tracking of the transducer shaft and electromagnetic tracking of the flexible transducer tip. A novel approach of modeling the poses of the transducer tip in relation to the transducer shaft allows us to reliably detect and significantly reduce electromagnetic tracking errors. For detecting errors of more than 5 mm, we achieved a sensitivity and specificity of 91% and 93%, respectively. Initial 3-D rms error of 6.91 mm were reduced to 3.15 mm.
Keywords :
biomedical ultrasonics; endoscopes; error correction; error detection; magneto-optical sensors; optical tracking; ultrasonic transducers; error correction; laparoscopic ultrasound transducers; magnetic tracking error correction; magneto-optical tracking; online error detection; transducer shaft; Electromagnetic fields; Electromagnetic interference; Error correction; Instruments; Laparoscopes; Magnetic susceptibility; Optical distortion; Shafts; Ultrasonic imaging; Ultrasonic transducers; Electromagnetic tracking; hybrid tracking; image-guided surgery; laparoscopic surgery; optical tracking; Calibration; Electromagnetic Fields; Equipment Failure; Humans; Laparoscopy; Online Systems; Optics and Photonics; Phantoms, Imaging; ROC Curve; Reproducibility of Results; Surgery, Computer-Assisted; Transducers; Ultrasonography;
fLanguage :
English
Journal_Title :
Medical Imaging, IEEE Transactions on
Publisher :
ieee
ISSN :
0278-0062
Type :
jour
DOI :
10.1109/TMI.2008.2008954
Filename :
4781567
Link To Document :
بازگشت